Suitable sensor and channel settings can be made here for each module
(slot) and each channel.
Click the slot numbers to select the measurement cards:
urement cards present in the PMX, gray = empty module slot (slot).
Example: force sensor at slot 3.1:
In the example above, module 3 is fitted with a PX455, and the first channel is
assigned to a force sensor (SG full bridge).
− The PMX amplifier channel is set to the full bridge sensor type, with the
measuring range of 4mV/V.
− Amplification is set to 1000 N, for a sensor sensitivity of 2mV/V.
− The filter type is Bessel, the filter frequency is set to 5 Hz (average to high
attenuation).
− The data are now changed in the PMX and are displayed by a floppy disk
symbol in the status line
− Press this symbol to store the settings power failsafe in the PMX (security
prompt).
orange = selected measurement card, blue = meas-
.
Quick Start
A3257−2.0 en/de/fr