HBM PMX Manuel D'emploi page 487

Masquer les pouces Voir aussi pour PMX:
Table des Matières

Publicité

Jeu de commandes
ROP
Syntax:
ROP p1, p2(x)
Parameter:
p1: binary representation of all outputs, 0...65535
p2: optional, binary representation of the selected outputs. If
omitted all 16 outputs are set to the state defined with p1.
Value of p2: 0...65535, default: 65535
Effect:
Sets the 16 possible digital outputs of the PMX device. The
lower 8 bits of p1 represent the 8 outputs of the first PX878.
The higher 8 bits represent the 8 outputs of the second
PX878.
The outputs are available even if no PX878 is plugged in. In
that case they are virtual. They can be set or read back, but
are not electrically present.
p2 defines the selected bits whose corresponding output is
set or cleared with p1. The outputs whose corresponding bits
are 0 in p2 are not affected.
Example:
ROP2, 32770(x)
0(y)
Output 8 (Bit 15) of the 2
1) of the 1
Only these two bits are selected with p2. All other outputs are
unchanged.
Note:
The outputs of the PMX device are defined by configurations
which are stored in (sub-) parameter sets which can be addi-
tionally switched. This command changes the configuration of
the selected outputs of the currently used sub-parameter set
in such a way that the output switches to the desired state. If
a parameter set is switched a previous ROP command is
overwritten. So be careful what you do!
A3257−2.0 en/de/fr
Set Digital Outputs
Digitalausgänge setzen
st
PX878 is set (outputs start counting with 1).
nd
PX878 is cleared and Output 2 (Bit
151

Publicité

Table des Matières
loading

Table des Matières