HBM PMX Manuel D'emploi page 94

Masquer les pouces Voir aussi pour PMX:
Table des Matières

Publicité

94
Default
Exception handling
- Lower threshold of capture range threshLow
- Capture outside (yes/no), capture outside interval
[threshLow, threshHigh]
- On entry into interval only (onEntryOnly) (yes/no)
Yes: The value is captured on entry into the interval (if
necessary, after expiration of delay) and held until the
next entry.
No: The output is permanently updated (if necessary,
after expiration of delay) for as long as the capture
criterion is met.
- Delay (delay) in ms.
Range 0.0 to 60000.0 ms (= 1 min); Resolution
0.1 ms
The value is only held if the capture criterion has per-
manently been active during this time period
- Invalid outside (invalidOutside) (yes/no)
The output value is marked as invalid for as long as
the capture criterion has not been met
- Initial value (initValue)
Is output in the event of a positive edge on the digital
init input
Upper threshold: 0.0
Lower threshold: 0.0
Capture outside: No
On entry into interval only: No
Delay: 0.0 ms
Outside invalid: No
Initial value: 0.0
If input in0 is marked as invalid, the capture function is
nonetheless carried through. The output is marked as in-
If control input is invalid, the capture function is not carried
through. The current output value is held.
valid on capture.
Quick Start