HBM PMX Manuel D'emploi page 128

Masquer les pouces Voir aussi pour PMX:
Table des Matières

Publicité

128
STP
Syntax:
STP(x)
Parameter:
none
Response:
none
Note:
Response depends on SRB command.
OMP
This command is used to position the read pointer in the system memory
(FIFO memory in which measured values are recorded). The user needs to be
careful. There is no error management!
Syntax:
OMP p1, p2 (x)
Parameter:
p1: -N,.,N , offset to move read pointer: -(max fifo lines
-1)...max fifo lines -1 max fifo lines from TSV - command
P2: 0,..,2 measurement rate group 3 asynchronous FIFO
If the measurement rate group (p2) is not specified, measurement rate group
0 is addressed.
Note:
Response depends on SRB command.
OMP?
Syntax:
OMP? p1(x)
Parameter:
p1: 0,..,2 measurement rate group
Response:
q1, q2 (y)
q1: available (readable) lines, current read pointer to current
write pointer
q2: data recording status
If the measurement rate group is not specified, measurement rate group 0 is
addressed.
Stop
Terminate measurement output and data acquisition
Output Measuring Pointer
Output Measuring Pointer Query
0 Data acquisition not running
1 Data acquisition running
Commmands
A3257−2.0 en/de/fr

Publicité

Table des Matières
loading

Table des Matières