148
Response:
0: command successfully executed
?: an error has occurred
For further error status use the commands EST? and
TED?100 and TED?101
Note:
Response depends on SRB command.
TED?
Syntax:
TED? p1(x)
Parameters:
P1
Effect
0
Reads the TEDS header (8 byte binary) from the TEDS transducer
q1: binary with "#" and blocklength (16Bit binary). There is no CR/LF at the
end of the binary TEDS data.
If more than one subchannel is selected (PCS/SPS) the data is separated by a
;
"
"
Virtual subchannels (channel 9) do not support TEDS (not physically present).
1
Reads the TEDS data from the memory of the amplifier.
q1: binary with "#" and blocklength (16Bit binary). There is no CR/LF at the
end of the binary TEDS data.
The blocklength depends on the TEDS (onewire) chip. E.g. 512 bytes. If more
than one subchannel is selected (PCS/SPS) the data is separated by a "
The minimal amount of bytes should be 31 (1 byte chks is removed from the
32 byte page)
SRB
Syntax:
SRB p1(x)
Parameter:
Effect:
There are two types of commands:
a.) Query commands (e.g. RMV?) are identified by a question
mark and generate output data regardless of the Response
behavior selected for the interface. It is not possible to stop
Transducer electronic datasheet Query
Select Response Behavior
Select the Response behaviour of the current interface
p1 Switch Response output on/
off
0
Switch off Response output
1
Switch on Response output
Commmands
;
"
A3257−2.0 en/de/fr