Holds the input value controlled via a digital input
- Digital Hold input (edge-controlled)
- Digital Reset input (only effective if post-triggering not
Value on hold out0
- Inverting of the Hold input (yes/no)
Yes: Hold if high -> low
No: Hold if low -> high
- Post-triggering possible (yes/no)
Yes: The output is updated with every edge of the
Hold input
No: The output is updated only with the first edge
after Reset
- Delay hold (delay) in ms.
Range 0.0 to 60000.0 ms (= 1 min); Resolution
0.1 ms
The value is only held if the Hold input has perman-
ently been active during this time period
Invert Hold: no Post-triggering possible: yes Delay hold:
If the input is marked as invalid (Invalid bit), the output is
also marked as invalid.
The Hold function is nonetheless carried through and out-
Captures an input value controlled by a measured value
(or a computed channel)
Capture criterion: If the control input is within (optionally
outside) the interval [threshLow, threshHigh].
If the capture criterion is met, input in0 is output to output
out0 (including status bit).
If the capture criterion is not met, the output is held at the
value captured last and, if necessary, cancelled.
- Control input (control)
- Digital Init input (controls initialization, high-active)
Captured value out0
- Upper threshold of capture range threshHigh
- Input in0
possible)
0.0 ms
put.
- Input in0