Commands
RMV?
Syntax:
RMV? p1 (x)
Parameter:
p1 Signal
Effect:
The RMV? command outputs the desired signal of the chan-
nels selected with PCS and SPS if possible. Not every chan-
nel type supports every signal type. If a channel is selected
that do not support the signal type passed in parameter p1,
2.0e20 will be displayed.
The signals Min, Max, Peak/Peak have to be "defined/cre-
ated" at top level before (see command "MSS"), if they
are to be displayed ! Otherwise 2.0e20 will be returned.
If p1 is omitted gross values will be returned.
Example:
PCS3,4(x) 'Select channels 3 and 4
0(y)
SPS1(x) 'Select subchannel 1 (of channels 3 and 4)
0(y)
RMV?214(x) 'Fetch measured values.
9.998,8.888(y)
Virtual subchannels (channel 9) do not support Min, Max, Peak/Peak subsi-
gnals.
A3257−2.0 en/de/fr
Read Current Measurement Value
Output the measuring data.
Signal
p1
214
Gross
204
Min
205
Max
217
Peak/Peak
131