Jeu de commandes
CAL
Calibrate amplifier, all selected channels (PCS / SPS). Implicitly unlocks
cal-option. The ACL setting is NOT changed ! Only supported for PX455 !
Syntax:
CAL(x)
Parameter:
none
Note:
With all CF­bridge amplifiers, this command triggers calibra­
tion. Measurement values "flicker" for several seconds.
Response depends on SRB command.
CAL?
Status of calibration procedure , all selected channels (PCS / SPS). Only
supported for PX455 !
Syntax:
CAL?(x)
Parameter:
none
Response:
q1(y)
A3257−2.0 en/de/fr
Calibrate, (enables and) starts cal procedure once
Status of Calibration procedure
0
Autocal not running
Autocal running
1
135