HBM PMX Manuel D'emploi page 130

Masquer les pouces Voir aussi pour PMX:
Table des Matières

Publicité

130
RMB?
This command is used to output the measured values recorded in the system
memory.
For the output, the character string "#0" (2 bytes) is placed at the beginning of
the measured values (only in the first line); as many values as available, or as
have been requested, can then follow. Each value has the size of 4 bytes and
"float" format.
If more measured values are requested than are actually available, the output
routine remains in wait condition until more measured values arrive. CR LF is
output once as a terminator after the last line. The output format is set by us-
ing the MBF command. As this is always followed by output, even if there are
less measured values available than have been requested, it is advisable, be-
fore using the RMB? command, to use the OMP? command to find out how
many measured value lines there actually are.
Syntax:
RMB? p1,p2,p3(x)
Parameter:
p1: number of measured value lines to be output
p2: output mode
6409 from the current reading pointer; move the reading
pointer p1 lines forward
p3: 0,..,2 measurement rate group
If the measurement rate group is not specified, measurement rate group 0 is
addressed
Read Measuring Buffer Query
Commmands
A3257−2.0 en/de/fr