CAP
Affects all selected channels (PCS / SPS)
Syntax:
CAPp1,p2,p3(x)
Parameter:
p1: point number (1 or 2)
p2: measurement signal (unit depending on amplifier), if no
input value, then the current measured value is adopted
p3: display value
Effect:
The input characteristic curve is defined by 2 points. The input
signal and associated display value must be entered for each
point. Virtual subchannels (channel 9) do not support calibra-
tion points. Setting parameters p1,p2 and p3 is allowed but
will be ignored !
Note:
Response depends on SRB command.
CAP?
Output of input (transducer) characteristic points of all selected channels
(PCS / SPS).
Syntax:
CAP?<p1>(x)
Parameter:
p1:
Response:
q1,q2,q3(y)
q1:
q2:
q3:
Virtual subchannels (channel 9) do not support calibration points.
Answer for p1=1: q2=0, q3=0.
Answer for p1=2: q2=100, q3=100.
Calibration Point
Input of transducer (input) characteristic points
Calibration Point Query
number of point (1 or 2)
point number (1 or 2)
measurement signal (unit depends on amplifier)
value in displayed units