Electrical connection
I
TEDS module
Supply for external
transducers
Fig. 7.5: PX401 plug terminal pin assignment : current source
External transducers are supplied with power via the PX401 master card
(OUT + and OUT −).
Important
The individual measurement channels on the PX401 measurement card are
not electrically isolated from each other. The PX401 measurement card has
electrical isolation from the basic device mains. The power supply for external
transducers is not electrically isolated and corresponds to the PMX device's
power supply.
A3257−2.0 en/de/fr
Cable
shield
+ I
TEDS
OUT +
OUT −
Housing
"
20 mA