12. SKEMA 6 dental operatory disinfecting and maintenance plan

WHEN
Before starting work
After each treatment
At the end of the work day
SKEMA 6
- OPERATING INSTRUCTION
PART
Condensate drain cock
Spray ducts
(if the dental unit is turned off
for more than 48 hours)
CATTANI surgical separator
Single-use protective barriers
Strumenti
Spray ducts
Instruments
Camera
Curing lamp
Surgical suction tubes
Flexible protection
Single-use barriers
Contaminated surfaces
Spray ducts
Bowl fi lter
Bowl
Surgical suction fi lter
Surgical suction tubes
Hydraulic saliva ejector
Flexible protection
Single-use barriers
Dental chair and dental unit
surfaces
WHAT TO DO
Drain condensate from air ducts.
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present).
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present).
Insert a tablet into each suction fi lter of VF
CONTROL PLUS.
Put single-use barriers and fl exible protection
on dental chair and dental unit.
Lubricate micromotor and turbine handpieces.
Flush water ducts of sprays (if the TIME
FLUSHING system is present).
Sterilize or disinfect the exterior.
Disinfect the exterior.
Disinfect the exterior.
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of STER 3
PLUS solution with each suction tube.
Sterilize the contaminated fl exible protection.
Replace the single-use barriers.
Disinfect with STER 1 PLUS.
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present).
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present).
Clean the fi lter in running water.
The content must be disposed separately.
Clean with commercial detergents designed
for ceramic materials.
Do not use harsh products or acids.
Check the fi lter and replace it if the suction
capacity is reduced (code 97461845).
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of STER 3
PLUS solution with each suction tube.
Clean the saliva ejector fi lter.
Sterilize the fl exible protection present.
Remove the single-use barriers used.
Clean surfaces with STER 1 PLUS.
SEE PARAGRAPH
See paragraph 9.2.
See paragraph 7.2.1.
See paragraph 7.2.
See paragraph 9.5.
/
See documentation enclosed
with the instrument.
See paragraph 7.6.
See documentation enclosed
with handpiece.
See paragraph 5.8.
See paragraph 5.7 or 5.7.1.
See paragraph 9.4.
/
/
/
See paragraph 7.2.1.
See paragraph 7.2.
See paragraph 7.1.
See paragraph 7.1.
See paragraph 9.3.
See paragraph 9.4.
See paragraph 6.4.
/
/
/
GB
EN
165