5.6.1.2. BOOST functioning mode

In addition to the ultrasonic vibration of the bit, a modulation with the amplitude of the generated power (approx. +40% of the set power) is performed.
The modulation creates a percussion effect on the bone thereby making incision easier.
Three different modulation times can be set depending on the thickness of the bone tissue to be treated.
Menu with surgical detaratarer device extracted but not active.
• When the left-hand multifunction button (M1) is pressed, one of the three
available working programmes is selected.
Each working programme stores the following data:
- maximum power generated,
- the quantity of physiological solution supplied by the peristaltic pump,
- set modulation (Bone)
- the power variation mode during functioning.
• By pressing the right-hand multifunction button (
ging the operational settings is accessed.
Operational settings menu.
With the instrument extracted but not active, when the right-hand multifunction
button (
) the submenu is accessed where the following icons regarding
the settable functions are to be found:
For setting the type of modulation
Method for varying the supplied power
• Press the MENU button to scroll through the icons until you select the
function required.
• Setting the type of modulation.
Press the INCREASE and/or DECREASE buttons to modify the modula-
tion depending on the thickness of the bone tissue to be treated:
- Bone I : hard bone tissue (8ms);
- Bone II : Medium bone tissue (16ms);
- Bone III : soft bone tissue (64ms).
• Supplied power variation mode.
Press the INCREASE and/or DECREASE buttons to select the rotation
speed variation mode:
- icon
: linear variation, proportional to the movement of the foot control
lever,
- icon
: ON/OFF variation that entails the supply of maximum power
set on the activation of the foot control lever.
• To confi rm the settings selected you just need to leave this submenu
pressing the ESC multifunction button.