Operation of command for dental chair movements ( 3 ).
These buttons move the dental chair as follows:
- Chair seat up
- Chair back up
- Chair seat down
- Chair back down
To stop movement, release the button.
NOTE: all the dental chair movement controls are locked when an
instruments is active or the AUTOSTERIL system is working.
Left-hand button operation ( 4 ).
• Holding down the button for at least 2 seconds with instrument extracted:
Chip-air control: sends a jet of air to the turbine or the micromotor.
Air delivery is activated by pressing the button; the jet of air is interrupted
when the button is released.
• Briefl y pressing the button with the instrument extracted:
Activation or deactivation of instrument spray.
WARNING!
A short acoustic signal warns of the switch.
When the LED ( 8 ) is on, it indicates operation with spray.
Right-hand button operation ( 5 ).
• Holding down the button for at least 2 seconds with instrument extracted:
Water Clean System control: sends a jet of running water to instruments
such as the turbine, the micromotor and the scaler for rinsing the spray
ducts. Water is delivered when pressing the button (5); when the button is
released, the jet of water is interrupted and a blast of air is automatically
activated to eliminate any residual drops of liquid in the instrument ducts.
• Briefl y pressing the button with the instrument extracted:
Activation or deactivation of instrument spray.
WARNING!
A short acoustic signal warns of the switch.
When the LEDs ( 8 ) are on, they indicate operation with spray.
Right lever operation ( 6 ).
NOTE: the lever functions only with the instruments in their rest
position.
For safety reasons, the selected function starts only after the switch has
been briefl y actuated and then released.
• Lever pushed down:
"Dental chair automatic return" program activated.
• Lever pulled up:
Dental chair program "B" start.
Left lever operation ( 7 ).
NOTE: the lever functions only with the instruments in their rest
position.
For safety reasons, the selected function starts only after the switch has
been briefl y actuated and then released.
• Lever pushed down:
"Rinse position" (PR) program activated
NOTE: when the switch is actuated the second time, the dental
chair reaches its work position.
• Lever pulled up:
Dental chair program "A" start.
Protection against liquid penetration.
The foot control is protected against liquid penetration.
Degree of protection: IPX1.
Cleaning.
Clean the foot control using a suitable product (see Paragraph 1.4).
NOTE: if the foot control slips on the fl oor, dust the slip-proof
rubber found under the base with a dry cloth.
112