6.2. Instruments on assistant's board

All the instruments provided on the assistant's board operate in the same
manner as those on the instrument board.
More specifi cally:
• Syringe, see paragraph 5.3.
• Turbine, see paragraph 5.4.
• Scaler, see paragraph 5.6.
• LEDA curing light, see paragraph 5.7.
• T-LED polymerising lamp, see paragraph 5.7.1.
• C-U2 camera, see paragraph 5.8.
NOTE: the turbine and scaler are equipped with cock [ f ] to adjust
the water spray found under the assistant's board. The amount of air spray
cannot be adjusted in the turbine.

6.3. Suction tubes

To adjust the suction power, use the slider [ a ] located on the tube
handpiece.
The SKEMA 6 comes with the V.D.S. system that allows the suction line to
be dried using an automatic stop delay (about 2 seconds).
Removing the suction tubes
ATTENTION!
Always wear gloves to prevent contact with infected material When
removing the suction tubes.
Remove the suction tubes from the conveyor fi ttings by turning and twisting
the tube fi tting. Detach the suction tubes from the holders by turning and
twisting the tube fi tting.
ATTENTION!
Never directly grasp the suction tube.
Cleaning the suction tubes.
As the dental units may be equipped with different suction systems (liquid
ring or wet, air) carefully follow the instructions provided by the suction
system manufacturer when disinfecting the system regarding the product
to be used, times and directions for use.
WARNING!
To clean the suction system, it is recommended to use STER 3 PLUS
(CEFLA S.C.) diluted as follows: 1 measuring cap (if purchased
package) or 2 measuring caps (if free package) in one litre of water
(6% solution).
Sterilization.
• Suction tube holders: water steam autoclave up to 135°C (2 bar), minimum
time: 15 min.
• Suction tubes: soak to cold sterilize.
ATTENTION!
Never use procedures in which the temperature goes over 55 °C with
the tubes.
Maintenance
Periodically lubricate the O-rings of the cannula holder terminals (see
Paragraph 9.4.) using S1-Protective Lubricant for O-Rings (CEFLA S.C.).
Note about biocompatibility
Only suction tubes supplied with the dental unit and there after original
replacement tubes can be used. The suction tubes must comply to the
standard ISO 10993-1 Biological evaluation of medical devices.
SKEMA 6
- OPERATING INSTRUCTION
GB
EN
143