Manual disinfecting cycle with hydrogen peroxide

With the S.S.S. system, you can execute a manual disinfection cycle of the
water ducts of all the instruments on the dentist's instrument board and the
syringe on the assistant board.
To disinfect, proceed as directed below:
A) Prepare the disinfectant:
Pour pure PEROXY Ag+ (or 3% oxygenated water) into the tank with
the orange band without diluting it.
B) Putting in the disinfectant:
• Replace the bottle of distilled water [ a ] with the one containing PEROXY
Ag+ (see paragraph 7.2.).
NOTE: make sure there is at least 300 cc of liquid present.
• Make certain the spray cocks [ d ] found at the bottom of the board
are opened (if they are not, either water does not come out or too little
water fl ows out).
• Press the key to fi ll the cup 5 (fi ve) consecutive times, thereby fi lling
5 cups with water. This step is very important as it ensures that all the
distilled water in the ducts is replaced with the disinfectant liquid.
• Withdraw the instruments one at a time and let the water run for at
least 2 minutes using the CHIP-WATER function of the foot control
for the dynamic instruments (see paragraph 5.2) and the water key
for the syringes.
NOTE: at this point, the ducts contain disinfectant liquid.
Put the instruments back in place.
C) Disinfectant contact time:
WARNING!
PEROXY Ag+ must be left to dwell in the ducts for at least 10 mi-
nutes, but not more than 30 minutes.
D) Rinsing the ducts:
• Replace the bottle containing PEROXY Ag+ with the one containing
distilled water (see paragraph 7.2).
• As during the previous step, take out the instruments one at a time a
let the water run at least 2 minutes using the CHIP-WATER function of
the foot control for the dynamic instruments (see paragraph 5.2) and
the water key for the syringes.
NOTE: at this point, the ducts contain distilled water again and
the dental unit is ready to be used.
WARNING!
• After disinfection, make sure that you close the tank containing PEROXY Ag+ (it will lose effectiveness if exposed to air).
• It is good practice to perform a disinfection cycle at least once a day, preferably at the end of the day.
• It is strongly unadvisable to let PEROXY Ag+ dwell in the ducts for more than 30 minutes.
PEROXY Ag+ storage.
For proper storage of PEROXY Ag+ follow the manufacturer's instructions printed on the package.
It is important to keep the package tightly closed and store it in a cool place at a temperature not exceeding 25°C.
WARNING!
Never leave PEROXY Ag+ or oxygenated water in the tank ( a ) for longer than one month.
If you are going to be absent from the surgery for long periods of time (holidays), completely empty out the tank ( a ) before leaving.