9.4. Surgical suction

The surgical suction system must be sanitized using a product suitable for
this purpose.
WARNING!
To clean the suction system, it is recommended to use STER 3 PLUS
(CEFLA S.C.) diluted as follows: 1 measuring cap (if purchased
package) or 2 measuring caps (if free package) in one litre of water
(6% solution).
At the end of each surgical procedure.
• Perform an automatic fl ushing cycle (if present) or suck in about half a liter
of the disinfecting solution with each of the suction tubes used.
• Sterilize the ends of the tube holder in autoclave with water steam at up to
135 °C (2 bar), minimum time 15 min.
At the end of each work day.
• Draw in 1 liter of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
• After rinsing with water, execute an automatic fl ushing cycle or aspirate
about ½ litre of sanitizing solution with each of the cannulas used.
WARNING!
Whatever sanitizing product you use, follow the instructions given
by the manufacturer.
NOTE: after these operations, it is advisable to aspirate only air in
order to dry the entire suction system (5 minutes).

Once a week

Remove the cannula body from its cord attachment and lubricate the O-rings
( o ) using S1-Protective Lubricant for O-Rings (CEFLA S.C.).
Once a year
Replace the suction tubes and ends of the tube holder.

9.5. CATTANI surgical separator

At the beginning of each working day.
Insert a tablet [ v ] into each fi lter [ d ] of VF CONTROL PLUS (CEFLA S.C.).
WARNING!
ALWAYS wear gloves when carrying out this operation to prevent any
contact with infected material.
At the end of each surgical procedure.
• Suck in about half a liter of solution prepared with the selected disinfectant
with each of the suction tubes used.
• Sterilize the ends of the tube holder in autoclave with water steam at up
to 135°C (2 bar), minimum time 15 min.
At the end of each work day.
• Draw in 1 liter of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
• Once rinsed with water, draw in approximately half a liter of the solution
prepared with the selected disinfectant with each of the suction tubes
used.
NOTE: after disinfecting, it is good practise to draw in air only to
dry the entire suction system (5 minutes).
Every 15 days.
• Clean the separator container and probes with a soft sponge and neutral
detergent.
• Clean the drain valve for the separator's container with the device provided
for this purpose.
Once a year.
• By technician: check the siphons and drains, check all the internal tubes
and plastic and rubber parts subject to wear.
Before leaving the surgery empty for a few days.
• Start the aspirator and run it 20 - 30 minutes without sucking in liquids.
The aspirator will dry itself completely. As a result, salt caused by moisture
and basic substances will not form, salt that may cause fan seizure and
motor blockage.