Right-hand button operation ( 5 ).
Operation:
• Holding down the button for at least 2 seconds with the instruments in
rest position:
Activation of the "Automatic dental chair return" program.
• Holding down the button for at least 2 seconds with instrument extracted:
Water Clean System control: sends a jet of running water to instruments
such as the turbine, the micromotor and the scaler for rinsing the spray
ducts.
Water is delivered when pressing the button (4); when the button is
released, the jet of water is interrupted and a blast of air is automatically
activated to eliminate any residual drops of liquid in the instrument ducts.
• Briefl y pressing the button with the instrument extracted:
Instrument spray activation or deactivation.
WARNING!
A short acoustic signal warns of the switch.
When the LED ( 6 ) is on, it indicates operation with spray.
Wireless version.
This foot control can also be supplied in wireless version (see Paragraph
5.2.4).
Protection against liquid penetration.
The foot control is protected against liquid penetration. Degree of protec-
tion: IPX1.
Cleaning.
Clean the foot control using a suitable product (see Paragraph 1.4).
NOTE: if the foot control slips on the fl oor, dust the slip-proof rubber
found under the base with a dry cloth.

5.2.3. "Power Pedal" foot control

Description of the
parts.
1 Handle.
2 Foot control.
3
Dental chair movements.
4 Chip-air control or activation/deactivation of instrument spray function.
5 Water Clean System control or activation/deactivation of instrument spray
function.
6 Program "B" retrieval or patient rinse position activation.
7 Program "A" retrieval or automatic dental chair return activation.
8 Warning LED for operation with spray.
Foot control operation ( 2 ).
With instrument removed
- Pushing the pedal ( a ), the instrument is started.
The instrument's rpm (or power) can be adjusted by varying the pressure
exerted on the foot control.
NOTE: the foot control adjusts the speed/power of the instrument
from the minimum to maximum value set from the instrument board.
- Release the foot control to stop instrument operation.
NOTE: with water spray, at the end of work a blast of air is au-
tomatically activated to eliminate any residual drops of liquid in the
instrument ducts.
With instrument in place
When the foot control is pressed, all automatic dental chair movements
are automatically blocked.
SKEMA 6
- OPERATING INSTRUCTION
GB
EN
111