Adjusting the luminous intensity
Turn the knob (e):
• Clockwise to increase the luminous intensity.
• Anticlockwise to reduce the luminous intensity.
NOTE for lamps with NO TOUCH sensor: You can also adjust
the luminous intensity by keeping your hand still on the sensor ( k ) (sen-
sitivity range 5 cm).
When the luminous intensity adjustment function is activated, the lamp
emits an acoustic signal:
• 1 short BEEP: increased luminous intensity.
• 2 short BEEPS: reduced luminous intensity.
When the maximum or minimum luminous intensity is reached, the lamp
emits 2 long BEEPS and the adjustment function is deactivated.
If the lamp is set to maximum, the adjustment function reduces the lumi-
nous intensity and stops when reaching the minimum or when you take
your hand off the sensor.
If the lamp is not set to maximum (any intermediate value), the adjustment
function always increases the luminous intensity and stops when reaching
the maximum or when you take your hand off the sensor.
Removing the handles
Remove the handle after pressing the release button ( o ).
NOTE: to facilitate this operation, we advise you to hold the optical
unit fi rm with the other hand.
Cleaning and disinfection
WARNING!
Carry out these operations with the lamp off and after letting it cool down.
For cleaning and disinfection, use soft, non-abrasive disposable paper (avoid
using recycled paper) or sterile gauze.
• Lamp arms and optical unit cover: clean the outer parts using STER 1 PLUS
(CEFLA S.C.) or a similar product.
• Transparent screen: clean using a soft cloth for glass soaked in a solution
of neutral soap and water.
WARNING!
Absolutely do not use products containing alcohol.
• Removable handles: the handles can be sterilized in a steam autoclave at
135°C (2 bar). Minimum time: 15 min.
WARNING!
• It is strictly prohibited to use abrasive substances and detergents
based on trichloroethylene, benzene, turpentine or solvents in ge-
neral on any part of the lamp.
• Do not spray the product used directly on the optical unit.
• Everything used for cleaning and disinfection must be thrown away
after use.
Maintenance
The VENUS PLUS-L operating lamp does not require any particular maintenance operations.
WARNING!
No parts can be repaired in the fi eld. Should any malfunction occur, please call Technical Service.
Troubleshooting.
• The lamp does not come on when fl icking the switch or sweeping your hand across the NO TOUCH sensor (if present).
Check that the power supply is properly connected, and if yes, call Technical Service.
• The luminous intensity does not change when acting on the knob or the NO TOUCH sensor (if present).
Call Technical Service.
• The luminous intensity has considerably reduced.
- Check the luminous intensity by acting on the adjustment knob.
- Clean the front screen using a soft cloth.
Should the luminous intensity not return to the initial values, call Technical Service.
154
154
154
GB
EN
SKEMA 6
- OPERATING INSTRUCTION