• Take the light out of its housing on the assistant's board or instrument
board.
The relative image appears on the console display to signal the instrument
has been withdrawn.
• Turn the front of the light and/or fi ber optic to the position most suitable
for curing (wand, gun or intermediate position).
• Use the MODE button to select the desired cycle as previously directed
(the selected cycle is always indicated by the illuminated LED).
NOTE: the curing light has a permanent memory therefore the last
cycle used will always be present the next time it is used.
• Place the fi ber optic in the position required for curing.
NOTE: the fi ber optic should be placed as close to the material to
be cured as possible without touching it.
• Press button START to start the cycle.
WARNING!
Operation: runs 2 consecutive cycles, rests 60 seconds.
NOTE: when a programmed cycle is activated, the LEDs (1, 2, 3,
B, R, L) indicate the time that elapses (in multiples of 5 seconds) and
turn off every 5 seconds of operation.
The curing light comes with a beep that BEEPS when the cycle starts,
BEEPS every 5 seconds of operation and lastly BEEPS twice at the end
of the work cycle.
• Allow light emission to stop by itself. However, it can be stopped at any
time by simply pressing the START button again.
WARNING!
• The curing light is equipped with a system that signals malfun-
ctions by illuminating the LEDS in different combinations (see next
paragraph).
• The curing light is equipped with a cut-out.
Operation with the VENUS PLUS-L LED operating lamp
If you have the VENUS PLUS-L LED operating lamp, you can activate an
automatic function that reduces the luminous intensity of the operating lamp
to a minimum when extracting the polymerising lamp instrument.
The following icons indicate whether the automatic function is active or
inactive:
Automatic function to reduce the luminous intensity NOT active
Automatic function to reduce the luminous intensity active
To activate/deactivate this function, press the multifunction button on the right
(
).
Indicators.
The following indicators are provided on the control console to signal curing light failure:
• LED 5 and LED 1, green, constantly on.
Lamp does not emit any light. Contact technical service department.
• LED 5 and LED 2, green, constantly on.
Instrument start up controller failure.
Contact technical service department.
• LED 5 and LED 3, green, constantly on.
Power supply too low.
Contact technical service department.
• LED 5 and LED 4 fl ash continuously.
Handpiece cut-out tripped. These LEDS will continue to fl ash until the light has cooled down enough (about 5 minutes) for it to be used again.
If the problems persists, contact the technical service department.
Maximum curable thickness.
The maximum curing thickness with single cycles is 3 millimeters (refer to the instructions of the composite material used as well).
WARNING!
This thickness must not be exceeded as the layer may not be completely cured.
Safety guidelines
WARNING!
• The LED is a class 2 source (Po < 1mW) in compliance with international standard IEC 60825-1. NEVER CONTINUOUSLY STARE INTO THE
LASER SOURCE.
The light emitted may cause eye damage in the event of direct radiation without eye protection.
Eye protection must always be worn when using the curing lamp and do not direct the light beam in eyes.
• The light emitted may damage soft tissues (oral cavity mucous, gums, skin).
Be extremely careful to direct the light precisely on the material to be cured.