13. SKEMA 6 dental operatory disinfecting and maintenance plan

WHEN
Before starting work
After each treatment
At the end of the work day
SKEMA 6
- OPERATING INSTRUCTION
PART
Condensate drain cock
Spray ducts
(if the dental unit is turned off
for more than 48 hours)
Surgical suction fi lters
Single-use protective barriers
Strumenti.
Spray ducts
Instruments
Camera
Surgical suction tubes
Flexible protection
Single-use barriers
Contaminated surfaces
Spray ducts
Bowl fi lter
Bowl.
Surgical suction fi lter
Surgical suction tubes.
Hydraulic saliva ejector
Flexible protection
Single-use barriers
Dental chair and dental unit
surfaces
WHAT TO DO
Drain condensate from air ducts
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present)
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present)
Place a VF CONTROL PLUS tablet inside
each fi lter
Put single-use barriers and fl exible protection
on dental chair and dental unit
Lubricate micromotor and turbine handpie-
ces
Flush water ducts of sprays (if the TIME FLU-
SHING system is present)
Sterilize or disinfect the outside
Disinfect the outside
Do not use harsh products or acids
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of STER 3
PLUS solution with each suction tube
Sterilize the contaminated fl exible protection
Replace the single-use barriers
Disinfect with STER 1 PLUS
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present)
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present)
Clean the fi lter in running water
The content must be disposed separately
Clean with commercial detergents designed
for ceramic materials.
Do not use harsh products or acids.
Check the fi lter and replace it if the suction
capacity is reduced (code 97290068)
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of STER 3
PLUS solution with each suction tube
Clean the saliva ejector fi lter
Sterilize the fl exible protection present
Remove the single-use barriers used
Clean surfaces with STER 1 PLUS
SEE PARAGRAPH
See paragraph 9.2
See paragraph 7.2.1
See paragraph 7.2
See paragraph 9.3
/
See documentation enclosed
with the instrument
See paragraph 7.6
See documentation enclosed
with handpiece
See paragraph 5.8
See paragraph 9.4
/
/
/
See paragraph 7.2.1
See paragraph 7.2
See paragraph 7.1
See paragraph 7.1
See paragraph 9.3
See paragraph 9.4
See paragraph 6.4
/
/
/
GB