Cleaning the instrument board
Clean the instrument board; use a suitable product approved against
bacteria and fungi and effective against viruses (including HBV and HIV);
see paragraph 1.4.
ATTENTION!
Do not use acids or harsh cleaning products.
NOTE RS version boards: The instrument holder [u] and cord
cover [z] are autoclavable at 135°.
NOTE CP version boards: the board's protection [ v ] can be
sterilized in autoclave at 135°.
Removable instrument cords
All the instruments have removable cords to ease cleaning.
Follow the directions given below to remove the instrument cords:
RS version boards
• Open the front of the board, after releasing it with the button [k] found behind
the console as shown in the fi gure.
• Loosen the plastic ring nuts and remove the cords.
CP version boards
Loosen the plastic ring nuts found underneath the board and then remove
the cords.
ATTENTION!
• Shut off the operatory unit before attempting to take off the cords.
• After shutting off the operatory unit, empty the syringe's ducts by
pressing and holding down the relative air and water buttons directly
on the bowl until water spray is no longer present.
• The cords of the TURBINE, MICROMOTOR and SCALER contain
water, therefore hold the end of the cord on the handpiece side over
the bowl When removing the cord.
• When putting a cord back on, make certain the contacts are perfectly
dry and the plastic ring nut is tight.
• Each cord may be remounted only in the position for the correspon-
ding instrument.
Clean the instrument board with a suitable product approved against
bacteria and fungi and effective against viruses (including HBV and HIV);
see paragraph 1.4.
ATTENTION!
The instrument cords are NOT suitable for autoclave or cold steri-
lization.
SKEMA 6
- OPERATING INSTRUCTION
GB
GB
85
85
85