OSD control panel.
This menu allows a number of handpiece C-U2 (see paragraph 5.8.1)
operating parameters to be edited. It appears directly on the screen.
Press fast selection button MIN to access the OSD control panel.
Press multifunction button ESC to quit the OSD control panel.
NOTE: The OSD control panel can also be accessed with the foot
control: place the handpiece in the work position and actuate the foot
control at least 2 – 3 seconds within the fi rst 10 seconds.
MyRay Grabber
This program allows the C-U2 PRO camera to be
to a PC / WORKSTATION.
A full description of how the MyRay Grabber works is given in the
instructions, provided in electronic format, enclosed with the C-U2 PRO
camera.
Disposable infection control sheaths
The camera can be a vehicle of cross-contamination between patients.
Always use disposable infection control sheaths (code 97901338) and
disinfect it on the outside after use every day.
The sheath (with white paper backing) is enclosed in two protective layers:
a transparent one with blue tab at the front and a paper one at the back.
Follow the directions below to install a new infection control sheath:
1. Insert the camera handpiece tip between the layer with white tab and
the rear paper backing. The lens, surrounded by the LEDS, must face
downwards towards the rear paper layer. Gently push the handpiece
to the end of the sheath.
2. Pull the blue tab removing the protective covers.
3. The dental camera is now protected and ready for use.
ATTENTION!
• Always make certain the handpiece is correctly inserted inside the
infection control sheath.
• Always change the disposable infection control sheath before using
the device on a new patient. This is a fundamental step in preventing
cross-contamination.
• Disposal: the disposable infection control sheaths are to be treated as
special waste materials (like surgical gloves).
Cleaning and disinfecting.
Clean the handpiece after each use (see paragraph 1.4.).
ATTENTION!
• The intraoral camera is not designed for cold sterilization by being
soaked, for example in solutions such as glutharaldeide or hydrogen
peroxide.
• All products must be used as directed by the manufacturer.
• All material used to clean and disinfect must be thrown away.
Maintenance and repairs.
The C-U2 PRO camera system does not require any particular maintenance.
In the event of malfunctions, please send back the complete handpiece.
ATTENTION!
There are no parts that can be repaired on site. In the event of a malfun-
ction, please contact an authorized dealer.
Returning parts.
• Please send back any defective devices in their original packaging. Do not
reuse damaged boxes.
• The device must be disinfected before being shipped to prevent cross-
contamination. Handpieces that have not been adequately cleaned and
disinfected will not be accepted
ATTENTION!
The sender shall be held responsible for any equipment damaged incurred du-
ring shipment regardless of Whether or not the devices are under warranty.
SKEMA 6
- OPERATING INSTRUCTION
setup when it is connected
GB
GB
121