9.5. CATTANI surgical separator

At the end of each surgical procedure.
• Suck in about half a liter of solution prepared with the selected disinfectant
with each of the suction tubes used.
• Sterilize the ends of the tube holder in autoclave with water steam at up
to 135°C (2 bar), minimum time 15 min.
At the end of each work day.
• Draw in 1 liter of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
• Once rinsed with water, draw in approximately half a liter of the solution
prepared with the selected disinfectant with each of the suction tubes
used.
NOTE: After disinfecting, it is good practise to draw in air only to
dry the entire suction system (5 minutes).
Every 15 days.
• Clean the separator container and probes with a soft sponge and neutral
detergent.
• Clean the drain valve for the separator's container with the device provided
for this purpose.
Once a year.
• By technician: check the siphons and drains, check all the internal tubes
and plastic and rubber parts subject to wear.
Before leaving the surgery empty for a few days.
• Start the aspirator and run it 20 - 30 minutes without sucking in liquids.
The aspirator will dry itself completely. As a result, salt caused by moisture
and basic substances will not form, salt that may cause fan seizure and
motor blockage.
How to remove the separator's container
ATTENTION!
Gloves must be worn when carrying out the following operation to
prevent contact with infected material.
• Move the dental chair fully up.
• Open the cover at the side of the hydrogroup [ a ] (see paragraph 7.7)
and turn the electrical box [ b ].
• Turn the electrical box [ b ] and hydrogen peroxide tank [ a ] (if
present).
• Completely empty the separator's container by pressing the timed button
[ c ] found on the cover.
• If present, remove the valve ( s ) for central systems.
• Turn and raise the container until it is detached from the drain pump
(k).
• Detach the container [ d ] from the cover [ f ] by pulling up the two side
elastic bands [ e ].
• Once cleaned, put the container [ d ] back in place greasing the drain
valve and o-ring with S1-Protection for o-rings.
• Put the electrical box [ b ] back in place and put the cover [ a ] back on
the side of the hydrogroup (see paragraph 7.7).
142