5.10. Peristaltic pump

This device allows saline solution to be supplied through a single-use
administration line without any contact.
This device is provided together with the micromotor and scaler.
NOTE:
• In order to use with the micromotor, contrangles with external cooling or for
hollow burs are required (Goldspeed R20-L type).
• In order to use with the scaler, handpiece SURGISON 2 is required.
Description of the symbols present on the device.
1) Material meets and exceeds the essential requirements of directive EEC
93/42 and subsequent changes.
2) CAUTION: PINCH HAZARD.
Do not put your fi ngers in rotating parts.
3) Material sterilized with ethylene oxide
4) Expiration date (yyyy-mm).
5) Single-use material.
6) Material identifi cation code
Putting into service
• Direct and put the IV drip pole [ a ] in place and hang the fl ask or bottle [ b ]
that contains the saline solution.
• Open the bag [ c ] and take out the sterile administration line.
WARNING!
Use sterile disposable gloves.
WARNING!
Check the condition of the packaging as well as the expiration date
of the administration line. Only CASTELLINI administration kits
guarantee proper trouble-free operation. These lines are sterile and
disposable, reusing them may put patients at risk of microbiological
contamination.
• Open the cover [ d ] of the peristaltic pump, turning it upwards.
• Attach the tube, being careful to place the part with the largest diameter
inside the pump's V seats. The pump rotates clockwise. Place the tube so
that the section that runs from the bag enters from the left side of the pump
(see fi gure).
• Close the cover [ d ]. If it does not close, open the cover again and check the
position of the tube.
WARNING!
Do not start the pump with the cover [ d ] open, fi nger pinch ha-
zard.
• Pierce the cap [ b ] of the bottle of saline solution with the outfl ow tip of the
administration line [ c ].
• Attach the administration line to the instrument cord using the plastic clips
provided in the sterile kit.
NOTE: use type A for the scaler's cord and type B for the micro-
motor's cord.
Operation
• In order to start the peristaltic pump, withdraw the desired instrument (micro-
motor or scaler) and press the relative button
NOTE: the icon ( D ) appears on the display and a BEEP is emitted
to signal the peristaltic pump has been started.
• If necessary, set the amount of saline solution delivered by the pump (see
paragraph below).
• At this point, the internal air and water supply is interrupted and the peristaltic
pump will automatically deliver saline solution when the instrument is star-
ted.
• To shut off the peristaltic pump, press the button again with the instrument
withdrawn.
.