5.6.1. SURGISON 2 surgical detartarer device

The Surgison 2 handpiece is intended exclusively for surgical bone ope-
rations in the odontoiatric fi eld. Any other use can cause serious injury to
the patient and the damage or breakage of the instrument.
Usage instructions.
CAUTION!
• Ordinary water from the mains must NOT be used for cooling the
operational fi eld. Make sure the cooling liquid used is of a suitable
sterile type.
• The SURGISON 2 handpiece must only be used with the bits supplied
with the instrument or other bits supplied by CASTELLINI S.p.A.
• Do not change the shape of the bit
• The instrument is supplied in a non-sterile state.
Use.
• Operational times: continuous operation.
• Put the instrument in the working position. The system automatically
detects the presence of a SURGISON 2 handpiece and displays the
relative menu.
• The maximum power generated can be changed by pressing the INCRE-
ASE and/or DECREASE buttons
NOTE: with the rapid selection buttons it is possible to automatically
set the following values:
- MAX button: 100% power
- MED button: 50% power
- MIN button: 25% power
• With the instrument in the working position, it is possible to choose two
different operational modes by pressing the MENU button:
1) NORMAL (see paragraph 5.6.1.1.):
- The bit vibrates subjected to a continuous (not modulated) power the
amplitude of which depends on the set value of the power.
2) BOOST (see paragraph 5.6.1.2):
- in addition to the ultrasonic vibration of the bit, a modulation in the
amplitude of the generated power (approx. +40% of the set power) is
performed. The modulation creates a percussion effect on the bone
thereby making incision easier.
- three different modulation times can be set depending on the thickness
of the bone tissue to be treated.
- There are 3 separate working programs.
• By pressing the
button for a longer time (at least two seconds), it is
possible to modify the amount of physiological solution supplied by the
peristaltic pump (see paragraph 5.10).
• To start up the instrument press the foot control (see paragraph 5.2).

5.6.1.1. Normal operational mode

The bit vibrates subject to a continuous power, the amplitude of which depends on the power value set.
Menu with surgical detaratarer device extracted but not active.
• When the right-hand multifunction button (
accessed to change the following operational settings:
Power variation mode
Press the INCREASE and/or DECREASE buttons to select the power
variation mode:
- icon
: linear variation, proportional to the movement of the foot
control lever,
- icon
: ON/OFF variation that entails the supply of maximum power
set on the activation of the foot control lever.
To confi rm the settings selected you just need to leave this sub menu
pressing the ESC multifunction button.
114
114
114
GB
GB
SKEMA 6
- OPERATING INSTRUCTION
) is pressed, the submenu is
100%
III
Bone
100%
100%