Error and Warning Messages
After pressing the START key: Red
and green tube status LEDs flash
slowly
Warning beep, 1x short
Device information
Green tube status LED lights up
All LEDs light up briefly and then
go out
Warning beep, 1x long
Suction does not automatically
shut off when instrument is
closed
Cause
The pump was started without an
inserted tube or the tube was not
detected by the pump, e.g. defec-
tive transponder.
Tube OK - irrigation can be started
Self-test concluded successfully.
The pump is ready
Leakage in tube system or defec-
tive vacuum suction
Error and Warning Messages
Troubleshooting
Insert a new tube into pump before starting.
Check the tube system and suction container for
leaks. If there is no leakage, suction is defective and
must be switched off. It is possible to continue using
the device without integrated vacuum pump. In this
case, use external suction. After the surgical proce-
dure, have the device checked by an authorized ser-
vice technician.