WARNING!
Filter
The vacuum tube with integrated filter is designed for max. 30 days. Replace the
vacuum tube if it is obviously contaminated. The filter prevents that body fluids
enter the interior of the device. Please note that the filter may reduce the suction
capacity.
WARNING!
The vacuum tube sets for this device contain diethylhexylphthalate (DEHP),
which is classified as toxic to reproduction according to the EU Directive 1272/
2008/EEC on Classification, Labeling and Packaging of Dangerous Substances.
DEHP may impair fertility and may cause harm to the unborn child. Therefore,
this product must not be used for unauthorized applications. When applied
within the intended use, the potential risk to pregnant or breastfeeding women
as well as to children resulting from the DEHP contained in this product is not
critical.
WARNING!
Cleaning of reusable tube set
The reusable tube set (NOT FOR SALE IN USA) is made of silicone, polysulfone
(PSU), and stainless steel. For cleaning and disinfection, use only pH-neutral or
slightly alkaline cleaning agents (e.g. neodisher MediClean 2.0 %), disinfectants
(e.g. Lysetol V 8 %), and drying and rinsing agents approved for the tube set ma-
terials.
Using unsuitable agents (e.g. neodisher MediKlar rinsing agent) may damage
the tube system and especially the PSU connectors.
WARNING!
Remove the membrane before cleaning.
WARNING!
Please check the reusable tube (NOT FOR SALE IN THE USA) for any signs of dam-
age after sterilization and before use. Never use a tube showing signs of dam-
age, especially brittleness and perforation.
WARNING!
Make sure to connect tubes as indicated by the matching connectors.
WARNING!
The day tube remains connected to the device after each surgery. The patient
tube must be disposed of immediately after surgery. A new sterile cap (included
with the patient tube) must be screwed onto the hygiene connector at the day
tube after each surgery. This sterile protector remains connected to the day tube
until the next surgery.
WARNING!
Do not touch
Avoid touching the roller wheel. Risk of injury!