Operating the Device
en
28
depending on the volume of the secretion container.
NOTE!
Only use suction containers with overflow protection.
5.11

Replacing the Suction Container

CAUTION!
Full suction containers must be replaced immediately without stopping surgery.
1. Stop the suction (see chapter 5.11 "Replacing the Suction Container").
Press the suction START/STOP key (LED goes out).
2. Actuate the trumpet valve or open the stopcock to vent the system.
3. Remove tubes from filled suction container.
4. Reattach tubes to empty suction container.
Close stopcock if using instrument with stopcock.
5. Press the suction START/STOP key (LED lights up).
6. Starting suction:
Operate the trumpet valve or open the stopcock.
5.12

Removing a Vacuum Tube

1. Remove the irrigation tube (4) from the instrument (5) (see Fig. 5-14 "Suc-
tion system with suction-/Irrigation instrument" or Fig. 5-15 "Suction system
with standard instrument").
2. Remove the suction container (3) from the irrigation tube (4).
3. Remove the vacuum tube (1) from the suction container (3).
4. Remove the hygiene filter from the vacuum tube (1).
5. Remove the vacuum tube from the device.