General Information; Device Description; Intended Use And Contraindications; Intended Use - WORLD OF MEDICINE PP110 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 121
3

General Information

3.1

Device Description

The device PP110 is a pump providing an irrigation and suction function for lap-
aroscopy.
The device itself is non-invasive and designed for use in multi-unit racks (non-
sterile area). It pumps medical sterile irrigation fluid through a sterile tube. This
flushes the corresponding fluids into the body cavity. The device can be used with
electrolyte-free media (such as glycine 1.5 % or sorbitol 3.0 %) and with isotonic
electrolyte-containing media (such as saline 0.9 % or Ringer's Lactate solution).
The device is equipped with a vacuum suction function used to suction off secret-
ed fluids. The secretion liquid is sucked through a tube into the container. The de-
vice functions only with the tube sets described in the accessory list (see chapter
12 "Accessory list").
3.2

Intended Use and Contraindications

3.2.1

Intended Use

WARNING!
To be used only with appropriate training
The device may only be used by surgeons or operating room personnel with the
necessary training in the corresponding indication.
The device PP110 is a suction and irrigation pump intended for use during diag-
nostic and/or therapeutic laparoscopic procedures to irrigate fluid into and re-
move fluid from the abdominal cavity.
3.2.2

Contraindications

Do not use this device PP110 for procedures where a specific pressure has to be
preset and reached such as a hysteroscopy or an arthroscopy.
3.3

General Warnings and Precautions

3.3.1

General Warnings

WARNING!
Technique and procedures
Only the physician can evaluate the clinical factors involved with each patient
and determine if the use of this device is indicated. The physician must deter-
mine the specific technique and procedure that will accomplish the desired clin-
ical effect.
WARNING!
Original accessories
For your own safety and that of your patient, use only original accessories.
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not operate the device in the vicinity of ex-
plosive anesthetic gases and not in the vicinity of oxygen-enriched environ-
ments.
General Information
Intended use
Contraindications
en
5

Publicité

Table des Matières