1. Connect the instrument.
Close stopcock if using instrument with stopcock.
2. Press the irrigation START/STOP key. The irrigation LED lights up.
3. Operate the trumpet valve or open the stopcock. Make sure tube set and in-
strument are completely filled with fluid. Remove all air from tube and instru-
ment.
4. Release the trumpet valve or close the instrument stopcock.
7.3

Suction Test

1. Close stopcock if using instrument with stopcock.
2. Press the suction START/STOP key (LED lights up).
The device generates a negative pressure inside of the suction system. The
vacuum pump is deactivated once a negative pressure of max. –60 kpa is
reached.
CAUTION!
The suction system has a leak if the vacuum pump does not stop. Check the suc-
tion system for leaks and repeat function test. Do not use device, if vacuum
pump is not operating or not operating correctly.
3. Open the instrument. Suction starts. The vacuum pump is automatically re-
activated if the min. negative pressure is exceeded.
4. Press the suction START/STOP key again (LED goes out).
The function test is complete. The device is tested and ready for surgery.
WARNING!
Device defect
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects and damage of the
power plug and power cable.
Functional Test
en
31