Care Of The Reusable Tube Set (Not For Sale In Usa); General Notes - WORLD OF MEDICINE PP110 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 121
1. Switch device off.
2. Disconnect device from power supply by pulling the mains plug from the
mains socket.
3. Remove power connection cable from mains socket.
4. The fuse holder is located next to the mains socket.
5. Remove fuse holder as depicted in Fig. 9-1 "Opening the fuse holder".
6. A Undo the latch of the fuse holder with a small screwdriver.
7. B Remove the fuse holder.
8. C Check fuse.
9. Insert a new fuse. Use only the specified type of fuse (see chapter 11 "Techni-
cal Data").
10. Insert the fuse holder until it can be heard snapping into place.
9.5

Care of the Reusable Tube Set (NOT FOR SALE IN USA)

All reusable inflow tubes must be cleaned, disinfected, and sterilized before use.
This applies in particular to the first use after delivery since the product is deliv-
ered non-sterile. Effective cleaning and disinfection are an essential prerequisite
for effective sterilization.
The device PP110 can be used with a reusable tube set listed as an accessory (see
chapter 12 "Accessory list"). Please comply with the following information when
using a reusable tube set.
9.5.1

General Notes

WARNING!
Reprocessing
Only reprocess after training has been completed. The reusable tube set may
only be reprocessed by trained personnel.
Within the scope of your responsibility for the sterility of the product in the ap-
plication of material compatibility, please note the following:
• Only cleaning/disinfection and sterilization methods sufficiently validated and
specifically designed for the tube set are to be used.
• The equipment used (washer-disinfector, sterilizer) must be regularly main-
Care and Maintenance
Fig. 9-1
Opening the fuse holder
en
35