CAUTION!
Position of the user
To avoid a malfunction, the user must be positioned correctly
• within a display viewing angle of ±50° to operate the device
• up to 2 m/6.5 ft from the device front for monitoring the actual values
CAUTION!
Disconnection from the power supply is only guaranteed if the mains plug is
pulled from the mains wall socket.
CAUTION!
Mains connection
• Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause er-
rors and malfunctions and may destroy the device.
• Make sure the connection data and technical specifications of the power sup-
ply comply with DIN VDE or national requirements. The mains connection ca-
ble may be plugged only into a properly installed, grounded safety wall socket
(shockproof socket) (see DIN VDE 0100-710).
• Read the device label located in rear of device (type plate) to determine the op-
erating voltage of the device.
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 and
IEC 60320-C13. Grounding will only be reliable if the equipment is connected to
a corresponding hospital grade socket.
Integrate the device into the potential equalization system as specified by local
safety rules and regulations.
Initial Device Startup
Mains connection
Grounding contact
Only for U.S. operators
Potential equalization