Guidelines And Manufacturer's Statement/Electromagnetic Interference Immunity - WORLD OF MEDICINE PP110 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 121
Electromagnetic Compatibility
en
Electromagnetic interference
immunity tests
Electrostatic discharges (ESD)
IEC 61000-4-2
Electrical fast transients/bursts
according to IEC 61000-4-4
Surges according to
IEC 61000-4-5
Blackouts, brownouts, and fluc-
tuations of the power supply
according to IEC 61000-4-11
Supply frequency (50/60 Hz)
magnetic field IEC 61000-4-8
Conducted RF IEC 61000-4-6
Transients RF IEC 61000-4-3
50
13.3
Guidelines and Manufacturer's Statement/Electromagnetic In-
terference Immunity
The PP110 is intended for use in an electromagnetic environment as described
below. The user/operator of the PP110 must make sure the device is operated
within such an environment.
IEC 60601 test level
Compliance levels
± 8 kV contact
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 15 kV air
± 2 kV for power lines
± 2 kV for power lines
Modulation 100 kHz
Modulation 100 kHz
± 1 kV line(s) to line(s)
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 2 kV line(s) to earth
0 % UT; 0.5 cycle
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
At 0°, 45°, 90°, 135°,
180°, 225°,270° and
180°, 225°,270° and
315°
315°
0 % UT; 1 cycle and 70 %
0 % UT; 1 cycle and 70 %
UT; 25/30 cycles
UT; 25/30 cycles
Single phase: at 0°
Single phase: at 0°
0 % UT; 250/300 cycles
0 % UT; 250/300 cycles
30 A/m
30 A/m
3 Vrms 0.15 MHz to
3 Vrms 0.15 MHz to
80 MHz
80 MHz
6 Vrms in ISM bands
6 Vrms in ISM bands
between 0,15 MHz and
between 0,15 MHz and
80 MHz 80 % AM at
80 MHz 80 % AM at
1 kHz
1 kHz
3 V/m 80 MHz to
3 V/m 80 MHz to
2.7 GHz
2.7 GHz
80 % AM by 1 kHz
80 % AM by 1 kHz
Electromagnetic environment guide-
lines
Floors should be made from wood or
concrete or covered with ceramic tiles.
If the floor covering consists of syn-
thetic material, the relative humidity
should be at least 30 %.
The quality of the supply voltage
should be the same as the voltage of a
typical business or hospital environ-
ment.
The quality of the supply voltage
should be the same as the voltage of a
typical business or hospital environ-
ment.
Magnetic fields of the mains power
frequency should comply with the typ-
ical values of business and hospital
environments.

Publicité

Table des Matières
loading

Table des Matières