Error and Warning Messages

en
Error and Warning Messages
Device error
All LEDs flash rapidly
Short warning beeps, 1x or 2x, are
continuously repeated
All LEDs flash rapidly
Short 4 warning beeps are contin-
uously repeated
All LEDs flash rapidly
Short 5 warning beeps are contin-
uously repeated
All LEDs flash rapidly
Short 7x warning beeps are con-
tinuously repeated
All LEDs flash rapidly
Short 10x warning beeps are con-
tinuously repeated
Suction does not start
Suction status LED flashes slowly
Short 3x warning beeps repeated
3 times
Tube warning
Irrigation cannot be started
Red and green tube status LEDs
flash slowly
Irrigation and suction status LEDs
light up
Warning beeps, 3x short beeps
The red tube status LED lights up
after inserting the tube
Warning beeps, 3x short beeps
The red LED flashes slowly after
pressing the start key while a
tube is inserted
Warning beeps, 3x short beeps
Irrigation cannot be started
Roller wheel stops
Red and green tube status LEDs
light up
Warning beeps, 3x short beeps
Green tube status LED is flashing
Warning beep, 1x short
52
14
Error and Warning Messages
Cause
Electronics defective
Sensor errors: Impermissible devia-
tion or error in the electronics
components measuring the pres-
sure. Pressure on sensor while no
tube is inserted.
Motor error: Defective motor con-
trol. The motor error warning can
also occur during the operation of
the device due to a jammed roller
wheel.
RFID error: Errors in the transpon-
der electronics
Calibration error: The device is not
calibrated properly.
Vacuum pump defective
The pump was switched on while a
tube was inserted.
The tube has no remaining uses
left or the tube is not approved for
this pump.
The tube has no remaining uses
left or the tube is not approved for
this pump. Start key was pressed.
The tube was detected by the tran-
sponder but the roller tube is not
correctly tensioned over the roller
wheel. Then the pump was started.
Last cycle for reusable tube After
the surgical procedure is finished,
the inserted tube cannot be used
any longer.
Troubleshooting
Turn the device off and back on after approx. 10 sec-
onds. If the error message is displayed again, the
device may not be used any longer. Make sure the
device can no longer be operated until a qualified ser-
vice technician conducts the appropriate tests and
repairs.
Turn the device off and back on after approx. 10 sec-
onds. If the error message is displayed again, the
device can be used without the integrated vacuum
pump. In this case, use external suction. After the sur-
gical procedure, have the device checked by an autho-
rized service technician.
Remove tube. Reinsert tube after a successful device
test. A tube may not be inserted while starting.
Reinsert the tube. If the tube is not detected again,
remove tube and insert a new one into the pump.
Reinsert the tube. If the tube is not detected again,
remove tube and insert a new one into the pump.
Remove tube and reinsert. Check the correct fit of the
tube on the left side of the roller wheel. The roller
wheel must be tensioned across the roller wheel.
Make sure a new tube is available for the subsequent
surgical procedure.