Workflow Automated Reprocessing
Workflow Manual Reprocessing
9.5.2

Reprocessing

9.5.2.1
Preparation
WARNING!
Membrane
Remove the membrane before cleaning.
1. Before cleaning: Carefully remove the membrane (2) from the pressure
chamber (1) (see Fig. 9-2 "Pressure chamber membrane and its position").
This is done by pulling the strap of the membrane (6) towards the top. Make
sure not to damage the membrane. Damaged membranes must be replaced.
2. Before starting sterilization, the membrane must be reattached to the pres-
sure chamber according to the instructions in chapter 9.5.2.6 "Reinstalling
the Membrane".
Care and Maintenance
Removing the membrane from the product
Fig. 9-2
Pressure chamber membrane
and its position
Pressure chamber
(1)
Membrane
(2)
(3)
Lip of the membrane
Groove
(4)
Notch
(5)
Strap
(6)
Spikes with protective caps
(7)
Tube clamps
(8)