Initial Device Startup
en
Setting up
12
4.2

Setting up and Connecting the Device

WARNING!
Always hold on to the power plug when disconnecting the device from the pow-
er supply. Never pull on the cable itself.
NOTE!
Locations
The device may be used only in the hospital or operating rooms.
Place the device on a flat surface free of vibration located in a dry environment.
The ambient temperature and humidity must meet the requirements mentioned
in chapter 11 "Technical Data" .
CAUTION!
ME systems (Medical Electrical System)
The medical electrical device is suitable for integration in ME equipment systems
(see chapter 13 "Electromagnetic Compatibility"). Operation of the ME device in
vicinity of non-ME devices may result in voiding the intended use of the ME de-
vice.
CAUTION!
ME Device in Rack
The ME device may not be used when stacked or stored directly adjacent to or
with other devices. If it should be necessary to operate the device close to
stacked with other devices, the ME device or ME system should be monitored to
ensure it works properly as configured.
CAUTION!
Device Setup
• Equipment should be positioned such that power cord can be easily discon-
nected.
• Position the device in such a way that it is easy to use and switch off and on.
• Place the device outside the sterile area.
CAUTION!
Ventilation of the device
• Avoid device overheating.
• Ensure free air circulation especially to the bottom and rear of the device (rear
panel distance of at least 10 cm).
CAUTION!
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.