Operating the Device
Remove the tube set
en
Connect the instrument
Insert tube set
(8)
(7)
Fig. 5-8
Inserting the roller tube set
22
1. To be carried out by non-sterile technician:
Open the outer packaging of the tube set and let you remove the inner tube
packaging by sterile personnel and open.
2. To be carried out by sterile technician:
Keep the instrument in the sterile area and hand the tube end with the
spikes (2) to the non-sterile technician.
Open the trumpet valve (12) of the suction and irrigation instrument.
3. To be carried out by non-sterile technician:
Before inserting, close both tube clamps (3) (Fig. 5-3 "Irrigation tube set el-
ements") on the bags.
Switch the device on. The power status LED on the ON switch lights up. Make
sure all tube sets are removed from the pump before switching the device
(12)
on. A long beep is emitted to signal the successful completion of the device
test and the tube can now be inserted.
Inserting the roller tube is depicted in Fig. 5-8 "Inserting the roller tube set".
When inserting the roller tube, make sure not to damage the membranes of
the pressure chamber. Insert the pressure chamber (8) only if chamber is
not pressurized.
Insert the unpressurized pressure chamber (8) carefully into the right notch
of the tube retainer (12) up to the stop.
Place the roller tube (7) around the roller wheel.
Insert the tube adapter into the left section of the tube retainer by tighten-
ing the tube.
Make sure the tube adapter is firmly snapped into place! Irrigation and suc-
tion tube must point away straight towards the front from the tube retainer.
Connect the suction tube with the suction container (17) (see Fig. 5-7 "In-
serting the suction/irrigation set").
5.7
Inserting a Day Patient Tube Set (NOT FOR SALE IN USA AND
CHINA)
The day patient tube set consists of two tube sections comprised of the day tube
and patient tube. The day tube establishes the tube connection between the flu-
id bags and the patient tube. The day tube is inserted into the tube retainer of the
device and not replaced for the duration of one day between surgeries and/or
surgical procedures. Only the patient tube (tube connection between day tube
and the instrument) must be replaced between surgical procedures. Efficiency is
increased with a day patient tube set if several surgeries are to be performed on
one day. Make sure to connect tubes as indicated by the matching connectors.
CAUTION!
Use only original day patient tube sets of the manufacturer. Other day patient
tube set are not compatible with the pump!

Publicité