Care and Maintenance

Special care is necessary when servicing, maintaining, and storing the device and
its accessories to maintain the functionality of the device and its accessories.
9.1

Cleaning the Device

1. Use the ON/OFF switch to turn the device off.
2. Remove the power cable.
3. Wipe the surface of the device with a soft cloth moistened with the surface
disinfectant based on alcohol (for example Meliseptol® rapid). The concentra-
tion of the used disinfectant depends on the information provided by the
manufacturer of the disinfectant. Make sure moisture does not enter the de-
vice.
CAUTION!
Cleaning the device
Do not sterilize the device.
9.2

Maintenance Intervals

WARNING!
Maintenance and calibration
Do not open the device. The device may not be opened or calibrated by the user.
Only authorized service technicians are permitted to repair, calibrate, or modify
the device or its equipment.
The manufacturer stipulates that qualified personnel or hospital technicians
must regularly test the device to assess its functionality and technical safety. This
inspection must be carried out once a year. The tests are described in chapter 10
"Annual Inspection".
Regular inspections will assist in early detection of possible malfunctions. This
helps preserve the device and increases its safety and service life.
NOTE!
Service or maintenance work may not be carried out during surgery.
9.3

Maintenance by Authorized Service Technician

An authorized service technician has to inspect and service the device at appro-
priate intervals to ensure the safety and functionality of the unit. The minimum
service interval is two years, depending on frequency and duration of use. If the
service interval is not maintained, the manufacturer does not assume any liabil-
ity for the functional safety of the device.
A sticker located on the rear panel of the device will remind you of the latest date
for the next service or maintenance check.
Authorized service technicians are only trained and certified by the manufactur-
er.
All of the service tasks, such as changes, modifications, repairs, calibrations, etc.
may be carried out only by the manufacturer or manufacturer-approved trained
and skilled technicians.
The manufacturer is not liable for the operational safety of the device if unau-
thorized persons conduct this maintenance or any other service tasks.
Unauthorized opening of the device and repairs performed by unauthorized per-
sonnel or third parties and/or changes or modifications release the manufactur-
Care and Maintenance
Manufacturer's specifications
Two-year maintenance interval
Authorized trained personnel
Unauthorized personnel
Liability