1.5. Potential adverse events
Potential adverse events in relation to flexible bronchoscopy (not exhaustive): Tachycardia/
bradycardia, hypotension, bleeding, bronchospasm/laryngospasm, cough, dyspnoea, sore
throat, apnoea, seizure, desaturation/hypoxemia, epistaxis, haemoptysis, pneumothorax,
aspiration pneumonia, pulmonary oedema, airway obstruction, reaction to drug or topical
anaesthesia, fever/infection, and respiratory/cardiac arrest.
1.6. General notes
If, during the use of this device or as a result of its use, a serious incident has occurred,
please report it to the manufacturer and to your national authority.
2. System description
The aScope 4 Broncho can be connected to the Ambu displaying unit. For information about
the Ambu displaying unit, please refer to the displaying unit Instructions for use.
2.1. System parts
Ambu® aScope™ 4 Broncho
– Single use device
aScope 4 Broncho Slim, aScope 4 Broncho Regular and aScope 4 Broncho Large are not
available in all countries. Please contact your local sales office.
Product Name
aScope 4 Broncho Slim 3.8/1.2
aScope 4 Broncho Regular 5.0/2.2
aScope 4 Broncho Large 5.8/2.8
2.2. Product compatibility
The aScope 4 Broncho Slim, Regular & Large have been designed to be used in conjunction with:
Displaying units
– Ambu aView
– Ambu aView 2 Advance
Endoscopic accessories
– Accessories with standard 6 % Introducer (Luer slip) and/or Luer Lock.
Endotracheal tubes (ETT) & Double lumen tubes (DLT)
– Tracheal tubes for use in anaesthetic and respiratory equipment in compliance with EN ISO 5361.
The aScope 4 family has been evaluated to be compatible wit the following endotracheal
tubes (ETT), double lumen tubes (DLT) and Endoscopic accessories (EA) sizes.
aScope 4 Broncho Slim
aScope 4 Broncho Regular
aScope 4 Broncho Large
Suctioning equipment
– Suction tube of diameters between 6.5 mm & 9.5 mm.
6
Part numbers
476001000 aScope 4 Broncho Slim 3.8/1.2
477001000 aScope 4 Broncho Regular 5.0/2.2
478001000 aScope 4 Broncho Large 5.8/2.8
Colour
Gray
Green
Orange
Minimum ETT
Inner diameter
5.0 mm
6.0 mm
7.0 mm
Outer diameter
[mm]
min 3.8; max 4.3
min 5.0; max 5.5
min 5.8; max 6.3
Minimum
EA minimum working
DLT size
channel width
35 Fr
Up to 1.2 mm
41 Fr
Up to 2.0 mm
Up to 2.6 mm
Inner diameter
[mm]
min 1.2
min 2.0
min 2.6

Publicité