Keep the power socket completely dry (see Figure 4).
For cleaning, Nanosonics recommends that the surfaces of the trophon2 device be cleaned in
accordance with organisational policies and following the detection of visible contamination. When cool,
wipe the chamber and outside surfaces of the trophon2 device with a cloth or wipe moistened with a
mild general purpose detergent, or soap and water solution, such as dish soap or general cleaners, until
all surfaces are visibly clean.
For disinfecting, Nanosonics recommends that the surfaces of the trophon2 device be disinfected in
accordance with organisational policies. Wipe all accessible surfaces of the trophon2 device with an
Isopropanol or Quaternary Ammonium (Quat) wipe.
SECTION E4: Transporting the trophon2 Device
NOTE: The procedure below is not necessary for transportation of the device, within a facility, using the
trophon Cart.
To transport the trophon2 device:
Purge the disinfectant. Do not insert a new disinfectant cartridge until the trophon2 device is
relocated.
Switch off the trophon2 device power switch and unplug from the mains.
Keep the trophon2 device upright at ALL times.
Pack using only Nanosonics-approved packaging.
SECTION E5: Disposal of the trophon2 Device
The trophon2 device is not biohazardous waste and shall be disposed in accordance with applicable
local Regulations. Dispose of trophon2 devices responsibly by contacting your Nanosonics
representative to advise the appropriate collection point for the recycling of electrical and electronic
equipment.
31
L03082 4.0 01/2023