C3.4 Closing the Chamber Door
The chamber door utilises a two-stage closure mechanism. Carefully close the chamber door to
the first click and do not force it shut. The door will automatically close further to seal and lock at
the start of the HLD cycle.
If the door is not properly closed, a screen message will prompt you to close the chamber door.
DO NOT force open or manually unlock the chamber door during a HLD cycle.
C3.5 Disinfecting the Probe
Ensure the probe is clean and dry. Start the cycle by either step 1 or 2 below:
1. Scanning the operator card on the AcuTrace reader (if AcuTrace and Log Operator Card is
enabled).
2. Pressing the start button on the screen (if AcuTrace and Log Operator Card is disabled).
The HLD cycle will take 7 minutes to complete, the cycle progress is indicated on the screen.
In the unlikely event that mist is escaping from the chamber, avoid direct contact with the
mist and remain at distance from the trophon2 device until completion of the operating
cycle and the mist is no longer visible. Contact your customer service representative.
Refer to Part D – Troubleshooting for further information.
C3.6 Removing the Probe
After completion of a successful HLD cycle, the ultrasound probe and chamber may have
surface temperatures up to 45 °C and 60 °C / 113 °F and 140 °F respectively. Take care not
to touch the chamber. The probe will feel warm to touch but is safe for handling and use
with gloves on.
1.
Wear gloves and follow the onscreen instructions.
2.
Using two hands, gently remove and wipe the probe with a dry, clean, single-use, low-lint or lint-
free cloth. Visually inspect the probe and remove any present disinfectant residue.
NOTE: Take care when removing the probe to ensure minimal contact and avoid contact with the
chamber. Remove the used trophon Chemical Indicator and verify the color change against the Color
Assessment Chart on the Chemical Indicator packaging. Discard the used Chemical Indicator.
a
Figure 15
26