Electrical Device
Use the power cable supplied with the trophon2 device, connect to an earthed power outlet with
the correct voltage and frequency as specified on the product and in Appendix 1: trophon2 Device
Technical Specifications N05000-2. Incorrect voltage can damage the product.
The trophon2 device must not be connected to the same circuit as critical patient or life support
equipment.
Spilled fluid can result in electrical shock. Avoid spilling fluids on or around the trophon2 device.
Do not immerse any parts of the trophon2 device in liquid.
Do not attempt to access the internal mechanics. This may result in electric shock.
Protective Wear and Spills
Wear clean disposable gloves throughout the complete high level disinfection (HLD) process
including but not limited to operating the trophon2 device and handling:
trophon NanoNebulant. Temporary bleaching and/or irritation of the skin may occur if
gloves are not worn.
Probes before and after a HLD cycle.
trophon Chemical Indicators before and after a HLD cycle.
Waste drawer when emptying or obtaining the Manual Door Lock key.
Wear appropriate personal protective equipment (PPE) when managing spills.
Never return spills to original cartridges for re-use.
Any serious incidents in relation to the trophon2 device or trophon NanoNebulant should be reported to
Nanosonics and / or the authority of your member state.
SECTION A2: Introduction to the trophon2 device
A2.1 Indications for Use
The trophon2 device is intended for the high-level disinfection (HLD) of non-lumened, reusable,
transiently invasive and non-invasive medical instruments/devices* e.g. devices that are intended for
use for imaging, diagnostic, ablation, coagulation and their accessories.
The trophon2 system consists of a multiple-use device, combined with a single-use disinfectant "trophon
NanoNebulant", delivered from a multi-dose cartridge.
The trophon2 device is suitable for use in general hospital and health care facilities by trained personnel.
The trophon NanoNebulant should be used with the following contact conditions:
Minimum Operational Cycle Time:
Minimum Concentration:
Minimum Disinfectant Dose:
Minimum Chamber Temperature:
The trophon2 device is NOT intended to reprocess single-use probes or instruments, or to pre-clean
medical instruments.
Chemical indicator use is required with every HLD cycle. Only the trophon Chemical Indicator product
is the approved chemical indicator for use with trophon2 device.
* The terms "ultrasound probe" and "probe" in the User Manual refer to approved medical instruments.
7 minutes
31.5%
1.0 g
56 °C
11