Figure 13
Curved Probes
The Integrated Probe Positioner (IPP) improves the positioning of approved curved probes in the
chamber of the trophon2 device. Refer to the trophon EPR and trophon2 Compatible Ultrasound Probes
List for further details of probes that can be used in the trophon2 device. Position the probe curve
towards the left of the trophon2 device's chamber. To engage the IPP, once the probe is inserted into
the probe clamp, push the left side of the IPP and slide to the right (see Figure 14). This should displace
the probe cable until the probe is no longer in contact with the chamber wall (Figure 13).
To release the IPP, depress the trigger (see Figure 15) and slide to the left.
NOTE: If the IPP is not required for the probe (as per the trophon EPR and trophon2 Compatible
Ultrasound Probes List), ensure that it has been completely released and fitted to the left.
Figure 14
25
L03082 4.0 01/2023