PART C – OPERATION
SECTION C1: Loading the trophon NanoNebulant Disinfectant Cartridge
The cartridge door opens automatically, DO NOT force it open.
Check expiry date of the trophon Nanonebulant. If expired, it cannot be used to run
disinfection cycles.
A trophon NanoNebulant disinfectant cartridge needs to be inserted into the trophon2 device before a
HLD cycle can commence.
Refer to the trophon NanoNebulant IFU for detailed instructions on how to scan and log, insert, or
remove disinfectant cartridges.
Inserting a Disinfectant Cartridge
The screen on the trophon2 device will automatically prompt users to scan (if AcuTrace is enabled) and
insert a new disinfectant cartridge, when required. Follow the onscreen instructions and refer to the
trophon NanoNebulant IFU.
Note: The trophon2 device will notify the user upon scanning, if the trophon NanoNebulant cartridge is
30 days or less from expiration
The trophon NanoNebulant will also expire 30 days after insertion into the trophon2 device, regardless
of the expiry date shown on the disinfectant label.
Once trophon NanoNebulant is expired a purge will be required. Refer to SECTION E2: Purge Cycle.
SECTION C2: Logging the trophon Chemical Indicators
No other chemical indicators are approved for use in the trophon2 device.
Check expiry date of the trophon Chemical Indicators. Expired Chemical Indicators cannot
be used to run disinfection cycles.
If the Log Chemical Indicators option is enabled, log the new batch of trophon Chemical Indicators:
1.
Select Menu
2.
Follow the onscreen instructions.
Note: When Log Chemical Indicators option is enabled, Chemical Indicators must be scanned at the
start of every new carton. As this is a manual process, it is important for users of the trophon2 device to
complete this step each time a new Chemical Indicator carton is opened. trophon Chemical Indicators
must be stored in their original packaging and not be shared across trophon devices outside of its original
packaging.
Note: The trophon2 device will notify the user during the first cycle of the day if the trophon Chemical
Indicators logged in the system are 30 days from expiry. Users will be reminded again 5 days before
expiry and everyday thereafter.
Do not use Chemical Indicators past their expiration date. Confirm Chemical Indicators
are within expiry before use.
Disinfection cycles will not run after the Chemical Indicators have expired. Log trophon
Chemical Indicators with valid expiry date to continue use of the trophon2 device.
Users will be reminded again on the first cycle, 1 day before expiry.
.
AcuTrace
Log Chemical Indicators.
L03082 4.0 01/2023
22