SECTION C3: Routine HLD Cycle
C3.1 Preparing the Probe
Wear gloves throughout the complete HLD process.
Clean and dry the probe and check carefully for probe defects prior to commencing the HLD process,
as per the probe manufacturer's instructions. Ensure that the probe has been thoroughly dried after
cleaning, with a low-lint or lint-free wipe and that no visible debris is present.
C3.2 Inserting the trophon Chemical Indicator
A whole trophon Chemical Indicator must be used for each disinfection cycle and may only be used
once. Refer to the trophon Chemical Indicator IFU.
C3.3 Positioning the Probe
1.
If AcuTrace is enabled, when prompted, scan the trophon AcuTrace Medical Instrument Tag
against the AcuTrace reader.
2.
Open the chamber door and load the probe and a Chemical Indicator.
3.
Two clamps (see Figure 10) hold the probe cable securely in the chamber.
4.
Wearing gloves, insert the probe into the trophon2 device by gently pulling the probe cable against
the cable clamp (see Figure 10a). Then carefully pull the cable upwards until the probe is
suspended in the correct location and the probe cable is held by the cable seal (see Figure 10b).
Refer to Figures 11–12 and note a correctly positioned probe in the trophon2 device.
Do not pull the probe down when loaded in the cable clamp.
a
Figure 10
23
L03082 4.0 01/2023
b