PART E – MAINTENANCE AND ROUTINE CARE
SECTION E1: Preventative Maintenance & Service
Your trophon2 device requires annual preventative maintenance and a major service at every 5,000
cycles.
Once the Service Due reminder screen message is displayed, please contact your customer service
representative to arrange a service. Contact details, if not provided by your distributor can be found on
the Nanosonics website.
The service options found: Menu
should only be used when advised by service personnel.
For more details, see SECTION G1: Service Schedule.
SECTION E2: Purge Cycle
The purge cycle removes and converts any remaining disinfectant from the trophon2 device into oxygen
and water.
E2.1 When to Run a Purge Cycle
Prior to transporting the trophon2 device, if it has previously been in use.
When an onscreen message states that your trophon2 device requires purging. This will occur
upon disinfectant expiry at 30 days after insertion. Follow the onscreen instructions to purge.
NOTE: Purging can be deferred until convenient on the day of expiry.
Initiate a purge cycle:
When the trophon2 device detects an error that requires a service call.
Before lifting or moving the trophon2 device.
For troubleshooting purposes when directed by service support only.
E2.2 How to Initiate a Purge Cycle
NOTE: Once the purge cycle has started it may be paused but it cannot be cancelled. Do not switch the
trophon2 device off during purging as this will restart the purge cycle. Do NOT attempt to open the
chamber or cartridge door during the purge cycle.
To initiate a purge cycle:
1.
Wear gloves and ensure that the waste drawer is empty and fully inserted into the trophon2 device
Refer to Figure 3 for location of the waste drawer.
2.
Select: Menu
The purge cycle will typically take less than 30 minutes.
3.
When purging is complete, put on gloves, empty the waste drawer and follow the onscreen
instructions.
4.
Load a new trophon NanoNebulant cartridge (unless transporting), follow the onscreen
instructions and refer to the trophon NanoNebulant IFU.
SECTION E3: Cleaning and Disinfecting
Do NOT submerge the trophon2 device, or pour liquids over it.
Keep the trophon2 device level and upright at all times.
Maintenance are used to troubleshoot device malfunctioning and
Maintenance
Purge. Follow the onscreen instructions.
L03082 4.0 01/2023
30