5.2
Alimentation externe
The Ascom 24V/DC power supply unit must be used for external power. The external power
requirement for the teleCARE IP system is 24 volt DC. The acceptable range of the power supply
voltage is 21.6V to 26.4V.
The Ascom 24V/DC power supply units are EN 60601-1-2 certified and "CE" marked. The input
mains power requirement is 230V/50-60HZ. The power supply unit must be plugged into an
"Essential" power outlet.
The power distribution is best organised with decentralised power supply units located as close as
possible to the rooms which they supply. This minimises the volts drop caused by the length of the
power supply wires and the load of the peripheral devices.
Note: In certain circumstances fuses in the 24V power supply are necessary.
5.2.1 Conformité VDE
Dans les installations conformes VDE, la prise secteur ne doit pas être utilisée. Les câbles
d'alimentation secteur doivent être vissés au secteur.
Dans les installations conformes VDE, l'alimentation électrique doit être raccordée à un onduleur
agréé conformément à la norme EN 60601-1. Tous les appareils raccordés au réseau du système
teleCARE IP doivent être connectés à l'onduleur et tous ces appareils connectés doivent être
conformes à la norme EN 60950-1. L'onduleur doit prendre en charge l'alimentation électrique du
système d'appel dans les 15 secondes qui suivent la panne d'alimentation et maintenir un
fonctionnement normal pendant au moins 1 heure. Toute panne d'alimentation électrique doit être
indiquée. Un dispositif de commutation omni-polaire doit être prévu pour l'arrêt du système.
AG
Figure 10. Conformité VDE 0834-1 requise
TD 92609FR