3.4

Operating principle

The LED lamp is very similar to a point light source.
The light is propagated through the optical cable and directed to the field
of view (operative field).
3.4.1

Brightness control

The brightness is adjusted in increments (5 % increments) via the bright-
ness control 4.
If a camera system with auto light quantity control is connected via the
communication interface cable OP942, the light intensity can be con-
trolled via the camera.
When the brightness control 4 is adjusted during active auto light quan-
tity control, the system is immediately switched to manual mode and the
light intensity is displayed in %.
3.4.2

Activation/standby mode

On pressing the brightness control with push function 4, the active light
source is put into standby mode. On pressing the brightness control with
push function 4 again, the light source goes back to the previously set
value.
If a camera system with auto light quantity control is connected via the
communication interface cable OP942, the LED light source can also be
activated or deactivated via the camera.
3.4.3

Multi optical cable socket

Various optical fiber cables from Aesculap/Storz, Olympus and Wolf with
an active diameter of 3.5 mm or 4.8 mm can be plugged into the multi-
optical cable socket 7.
3.4.4

Light source test unit with sterile adapter

The optical cable test unit can be used in the sterile field with a sterile
adapter OP941.
The optical cable test unit is used for testing the light transmission of the
4.8 mm optical fiber cable connected in each case.
The testing of 3.5 mm optical fiber cables causes false measurement
results.
3.4.5

Modes shown on the display

Display
STANDBY
100%
AUTO
4.

Preparation and setup

Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
When setting up and operating the product, adhere to
– The national installation and operator regulations
– National regulations on fire and explosion protection
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes (see IEC 60601-1/EN 60601-1 or the rele-
vant national standards).
Ensure that all devices operated in the vicinity meet their relevant EMC
requirements.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
Explanation
Standby
Device is switched on, LED is off.
Circular green light on the brightness
control is deactivated.
Light intensity % (0 % to 100 %)
Automatic light quantity control is active
Is only displayed if a corresponding cam-
era system is automatically controlling
the light quantity.
Optical cable is not plugged in or not
plugged in properly
Optical cable is being tested
Result of optical cable test:
Optical cable OK.
Result of optical cable test:
Optical cable should be replaced.
5