5.3

Optical cable test

If necessary, carry out an optical cable test:
Turn on the LED light source with the power ON 3 switch.
Plug the optical cable into the optical cable socket 7 until it audibly
snaps into place.
Screw sterile adapter OP941 onto the optical cable test unit 5.
Note
The sterile adapter OP941 consists of three parts. Of these, two bushes
serve the connection of the different optical fiber connections.
According to the optical fiber connection, unscrew or screw off the
bushes of the sterile adapter OP941. Keep the bushes.
Screw the distal end of the optical cable to be tested onto the sterile
adapter.
Activate the brightness control 4 to switch on the light.
Press the optical fiber unit 5 against the light source.
The results are displayed in the display.
Display
Explanation
Optical cable can still be used.
(100 % - 41 %)
Optical cable should be replaced.
(40 % - 0 %)
Note
The optical cable test takes place regardless of the selected light strength
(e.g. 80 %).
5.4

Safe operation

Risk of injury and/or malfunction!
Always carry out a function check prior to each
use of the product.
WARNING
Risk of injury due to application of the product
outside the field of view!
Apply the product only under visual control.
WARNING
5.4.1

Turning the device on and off

Turn on the LED light source with the power ON 3 switch.
The power ON indicator 2 light is on.
Turn off the LED light source with the power OFF 1 switch.
8
®
5.4.2
Brightness control
To adjust the brightness in increments (5 % increments), turn the

brightness control 4.

Turning it in a clockwise direction increases the light intensity.
Note
The brightness can also be automatically controlled using a corresponding
camera system.
5.4.3

Activation/standby mode

To activate the LED light source, press the brightness control 4.
To put the active LED light source (optical fiber cable is plugged in,
light is emitted) in standby mode, press the brightness control 4.
To take it out of Standby mode and put it back to the previously set
value, press the brightness control 4 again.
Note
The LED light source can also be activated/deactivated using a correspond-
ing camera system.
5.5

Ending the operation of the device

To end the operation of the device, press the power OFF switch 1.
Remove the optical cable from the sterile adapter 5.
To completely disconnect the device from the power supply, remove
the plug from the power cord connector 10 on the rear of the device.
6.

Disassembly

Sterile adapter OP941
Unscrew the two adapters 14 and 15 from the sterile adapter 16.
7.

Assembly

Sterile adapter OP941
Screw the two adapters 14 and 15 onto the sterile adapter 16.